2.5 mm Screw Craniomaxillofacial Specification, Uses & Sizes

  • 2.5 mm Screw Craniomaxillofacial available lengths are 8mm, 10mm, 12mm, 14mm, 16mm, 18mm, 20mm, 22mm, 24mm, 26mm, 28mm and 30mm.
  • Screws are made from pure Titanium.
  • Any additional length sizes of this screw will be made on demand.
  • This is Cross Head screw.
  • Instruments are available for this screw such as Bone Taps, Combined Drill & Tap Sleeve, Counter Sink, Depth Gauge, Drill Bits, Drill Guide, Drill Sleeve, Hollow Mill Screw Removal, Reverse Measuring Device, Screw Drivers and Screw Holding Forceps etc.

2.5 mm Screw Craniomaxillofacial

2.5 mm Screw Craniomaxillofacial Indications

2.5 mm Screw Craniomaxillofacial is intended for oral, maxillofacial surgery; trauma and reconstructive surgery, specifically for fractures of the subcondylar region of the mandible and fractures of the condylar basis region of the mandible.

2.5 mm Screw Craniomaxillofacial Contraindications

Screw is contraindicated for use in Acute or chronic, local or systemic infections, Allergy to implant material and Insufficient bone quality to secure implant.

Do you have any questions for 2.5 mm Screw Craniomaxillofacial?
Please fill this form we will get in touch with you shortly.

Name(Required)
This field is for validation purposes and should be left unchanged.

Surgical Technique

Expose and reduce fracture

After completing the preoperative plan, expose the fracture or osteotomy site. For trauma, reduce the fracture as required.

Select and prepare implant

Select the appropriate plate depending on the indication. Orient the plate so the topside is facing out.

Cut to length, if necessary. Mini and intermediate plates may be cut using the combination bending/cutting pliers. Large plates may be cut using the shortcut plate cutters.

Determine the appropriate screw type.

Contour the plate

Contour the plate to match the anatomy. An exact match is not required when using locking screws, because plate stability is not dependent on plate-to-bone contact when screws are locked.

Plates can be contoured using the combination bending/cutting pliers and combination bending pliers.

Position the plate

Place the plate over the fracture or osteotomy site. Use the plate holding forceps to secure the plate to the bone, if desired.

Drill the hole

Predrilling is recommended in complex fractures of the midface and in regions with thick cortical bone. Drill the first hole close to the fracture or osteotomy site.

Screw insertion

Insert the proper length locking or nonlocking screw through the plate and tighten until secure. Insert the second screw on the opposite side of the fracture or osteotomy site, and then all remaining screws, following the previously outlined procedure. Securely tighten all screws unless resection is to follow.

Resect the mandible

Once the plate is in place, remove the plate and screws, taking note of each screw’s placement.

Resect the mandible.

Replace the implants

Place the plate back onto the mandible in its original position. Reinsert each predetermined screw. Check all screws to ensure a secure fit in the plate.

Apply bone graft

A vascularized bone graft must be applied to constructs used in reconstructing the mandible.

More Products from Craniomaxillofacial Screws

Different Types of Screws including 2.5 mm Screw Craniomaxillofacial

Cancellous Screws

Cannulated Screws

  • 2.5 mm Headless Compression Screws Partially Thread
  • 3 mm Headless Compression Screws Partially Thread
  • 3.5 mm Headless Compression Screws Partially Thread
  • 4 mm Headless Compression Screws Partially Thread
  • 4.5 mm Headless Compression Screws Partially Thread
  • 5.5 mm Headless Compression Screws Partially Thread
  • 6.5 mm Headless Compression Screws Partially Thread
  • 7.5 mm Headless Compression Screw Partially Thread
  • 5 mm Interference Screw
  • 6 mm Interference Screw
  • 7 mm Interference Screw
  • 8 mm Interference Screw
  • 9 mm Interference Screw
  • 10 mm Interference Screw

Bone screws are the most commonly used orthopedic implants. There are many different types and sizes of screws for different types of bones. Most bone screws are made out of stainless steel or titanium alloys. The outer diameter, root diameter, and thread pitch and angle are important in determining screw mechanics.

In orthopedics, screws are typically described by their outer diameter, for example, a “2.5 mm Screw Craniomaxillofacial” has an outside diameter of 2.5 mm. The pitch of a screw is the linear distance travelled by a screw for one full turn of the screw. The screw advances by a distance equal to the distance between the threads with each full turn. Cortical screws have a lower pitch and therefore more number of threads. Cancellous bone screws have a greater depth of the screw to increase the surface area and therefore improve the purchase, as the bone is weaker.

Screws function by converting the tightening torque into internal tension in the screw and elastic reactions in the surrounding bone. This creates compression between the fracture fragments that the screw is holding together. Screw is typically inserted into holes drilled equal to the root diameter and are either self-tapping or are inserted tapped (threaded) holes. The torque to insert cortical bone screws can be high, so the screws must be properly inserted into the correct size drilled hole and designed to withstand insertion torque levels expected in cortical bone. Cancellous bone screws have large, deep threads that grip the spongy bone well. Because of the relatively low strength of the cancellous bone, failure of the screw itself during insertion is rare, but pull out can be an issue.

2.5 mm Screw Craniomaxillofacial Contraindications

Contraindications may be qualified or total, and need to be taken into consideration when evaluating the prognosis in each case. The physician’s education, training and professional judgement must be relied upon to choose the most appropriate device and treatment. Conditions presenting an increased risk of failure include:

  • Any active or suspected latent infection or marked local inflammation in or about the affected area.
  • Compromised vascularity that would inhibit adequate blood supply to the fracture or the operative site.
  • Bone stock compromised by disease, infection or prior implantation that can not provide adequate support and/or fixation of the devices.
  • Material sensitivity, documented or suspected.
  • Obesity. An overweight or obese patient can produce loads on the implant that can lead to failure of the fixation of the device or to
    failure of the device itself.
  • Patients having inadequate tissue coverage over the operative site.
  • Implant utilization that would interfere with anatomical structures or physiological performance.
  • Any mental or neuromuscular disorder which would create an unacceptable risk of fixation failure or complications in postoperative care.
  • Other medical or surgical conditions which would preclude the potential benefit of surgery.
  • All associated diseases which could endanger the function and success of the 2.5 mm Screw Craniomaxillofacial.

Warnings and Precautionary for 2.5 mm Screw Craniomaxillofacial

Before using 2.5 mm Screw Craniomaxillofacial, the surgeon and ancillary staff should study the safety information in these instructions, as well as any product-specific information in the product description, surgical procedures and/or brochures.

Screws are made from medical grade materials and are designed, constructed and produced with utmost care. These quality screw assure best working results provided they are used in the proper manner. Therefore, the following instructions for use and safety recommendations must be observed.

Improper use of  screw can lead to damage to the tissue, premature wear, destruction of the instruments and injury to the operator, patients or other persons.

It is vital for the operating surgeon to take an active role in the medical management of their patients. The surgeon should thoroughly understand all aspects of the surgical procedure and instruments including their limitations. Care in appropriate selection and proper use of surgical instruments is the responsibility of the surgeon and the surgical team. Adequate surgical training should be completed before use of implants.

Factors which could impair the success of the operation:

  • Allergies to implanted materials.
  • Localized bone tumours.
  • Osteoporosis or osteomalacia.
  • System disease and metabolic disturbances.
  • Alcohol and drug abuse.
  • Physical activities involving excessive shocks, whereby the implant is exposed to blows and/or excessive loading.
  • Patients who are mentally unable to understand and comply with the doctor’s instructions.
  • Poor general health.

Possible Adverse Effects

The following adverse effects are the most common resulting from implantation:

  • Loosening of the screw, which may result from cyclic loading of the fixation site and/or tissue reaction of the implant.
  • Early and late infection.
  • Further bone fracture resulting from unusual stress or weakened bone substance.
  • Temporary or chronic neural damage resulting from pressure or hematomata.
  • Wound hematomas and delayed wound healing.
  • Vascular disease including venal thrombosis, pulmonary embolism and cardiac arrest.
  • Heterotopic ossification.
  • Pain and discomfort due to presence of the 2.5 mm Screw Craniomaxillofacial.
  • Mechanical failure of the implant, including bending, loosening or breakage.
  • Migration of implant resulting in injury.

Preoperative Planning for 2.5 mm Screw Craniomaxillofacial

The operating planning is carried out following a thorough clinical evaluation of the patient, Also, x-rays must be taken to allow a clear indication of the bony anatomy and associated deformities. At the time of the operation, the corresponding implantation instruments in addition to a complete size of 2.5 mm Screw Craniomaxillofacial must be available.

The clinician should discuss with the patient the possible risks and complications associated with the use of implants. It is important to determine pre-operatively whether the patient is allergic to any of the implant materials. Also, the patient needs to be informed that the performance of the device cannot be guaranteed as complications can affect the life expectancy of the device.

2.5 mm Screw Craniomaxillofacial Precautions

  • Confirm functionality of instruments and check for wear during reprocessing. Replace worn or damaged instruments prior to use.
  • It is recommended to use the instruments identified for this screw.
  • Handle devices with care and dispose worn bone cutting instruments in a sharps container.
  • Always irrigate and apply suction for removal of debris potentially generated during implantation or removal.

2.5 mm Screw Craniomaxillofacial Warnings

  • 2.5 mm Screw Craniomaxillofacial can break during use (when subjected to excessive forces). While the surgeon must make the final decision on removal of the broken part based on associated risk in doing so, we recommend that whenever possible and practical for the individual patient, the broken part should be removed. Be aware that implants are not as strong as native bone. Implants subjected to substantial loads may fail.
  • Instruments, screws and cut plates may have sharp edges or moving joints that may pinch or tear user’s glove or skin.
  • Take care to remove all fragments that are not fixated during the surgery.
  • While the surgeon must make the final decision on implant removal, we recommend that whenever possible and practical for the individual patient, fixation devices should be removed once their service as an aid to healing is accomplished. Implant removal should be followed by adequate post-operative management to avoid refracture.

2.5 mm Screw Craniomaxillofacial General Adverse Events

As with all major surgical procedures, risks, side effects and adverse events can occur. While many possible reactions may occur, some of the most common include: Problems resulting from anesthesia and patient positioning (e.g. nausea, vomiting, dental injuries, neurological impairments, etc.), thrombosis, embolism, infection, nerve and/or tooth root damage or injury of other critical structures including blood vessels, excessive bleeding, damage to soft tissues incl. swelling, abnormal scar formation, functional impairment of the musculoskeletal system, pain, discomfort or abnormal sensation due to the presence of the device, allergy or hypersensitivity reactions, side effects associated with hardware prominence, loosening, bending, or breakage of the device, mal-union, non-union or delayed union which may lead to breakage of the implant, reoperation.