2.4 mm Locking L Plate Angle Specification, Uses, Sizes & Surgical Instruments.

2.4 mm Locking L Plate Angle

2.4 mm Locking L Plate Angle Specification

  • Plates available holes are 4, 5, 6, 7, 8 and 10.
  • Plates available for Left and Right both direction.
  • Available in both Titanium.
  • Plate has combi holes and round holes. Combi holes allow fixation with locking screws in the threaded section and cortex screws in the dynamic compression unit section for compression.
  • The shaft holes accept 2/2.4/2.7 mm locking screws in the threaded portion or 2/2.4/2.7 mm cortical screws  in the compression portion. Plate head accept 2/2.4/2.7 mm locking screws in the threaded portion.
  • Locking screws offer a fixed-angle construct to support the articular surface, reduce the need for bone graft, and obtain fixation in osteoporotic bone.
  • 2.4 mm Locking L Plate Angle allow implant placement to address the individual fracture pattern.
  • Limited-contact surface reduces bone-to-plate contact and helps to preserve the periosteal blood supply.
  • Choice of different lengths of plate eliminates the need to cut plates.
  • Polished surface and rounded edges minimize potential for tendon adhesion.
  • Low plate and screw profile minimizes potential for tendon and soft tissue irritation.
  • Smaller plates and screws address fracture fragments individually, with less overall implant bulk
  • locking plate increases construct stability, decreases risk of screw back-out and subsequent loss of reduction. It also reduces the need for precise anatomic plate contouring and minimizes the risk of stripped screw holes.
  • A complete Instruments Set is available for 2.4 mm Locking L Plate Angle. General Instruments are available for this plate such as Plate Bending Press, Plate Holding Forceps, Plate Bending Pliers, Bone Holding Forceps, Bone Elevators, Bone Cutter, Bone Nibbler, Depth Gauge, Sleeve, Screw Driver, Trocar Sleeve etc.

2.4 mm Locking L Plate Angle Uses

2.4 mm Locking L Plate Angle is intended for temporary fixation, correction or stabilization of bones in anatomical region of the hand.

Indications
–– Fracture fixation of the phalanges and metacarpals
–– Osteotomies
–– Arthrodeses
–– Replantations and reconstructions of phalanges and metacarpals, particularly in osteopenic bone

Instruments for 2.4 mm Locking L Plate Angle

Hand and Foot Instruments Set

A complete instruments set are also available for 2.4 mm Locking L Plate Angle. Instruments can be modified according to the customer’s requirement with minimum quantity required. All these instruments can be used several times.

We are keeping wide range of  instruments items in this set to ensures that Doctors get almost all required items during. Below is list of items of this set.

  • 2.4 mm Threaded Drill Guide for 1.8 mm Drill Bit
  • 2.7 mm Threaded Drill Guide for 2 mm Drill Bit
  • Counter Sink Copuling Shaft
  • Depth Gauge for 2 / 2.4 mm Screws
  • Depth Gauge for 2.7 mm Screws
  • Double Drill Sleeve (Universal Drill Sleeve) 1.8/2.4 mm
  • Double Drill Sleeve (Universal Drill Sleeve) 2.0/2.7 mm
  • Drill Bit 1.8 x 100 mm, Quick Coupling
  • Drill Bit 2 x 100 mm, Quick Coupling
  • Drill Bit 2.4 x 100 mm, Quick Coupling
  • Drill Bit 2.7 x 100 mm, Quick Coupling
  • Hohmann Recator 6 mm
  • Kirschner Wire 1.8 x 150 mm
  • Plate Bender
  • Periosteal Elevator Fiber Handle
  • Plate and Bone Holding Forceps Ratchet Lock-Type1
  • Plate and Bone Holding Forceps Ratchet Lock-Type2
  • Reduction Forcep Pointed Tip Ratchet Lock
  • Reduction Forcep Serrated Tip Ratchet Lock
  • Screw Driver Hexagonal, Holding Sleeve, for 2.7 mm Locking Screws
  • Screw Driver Torque for 2.4/2.7 mm Locking Screws
  • Screw Driver Quick Coupling for 2.7 mm Locking Screws
  • Sharp Hook
  • T Handle Quick Coupling
  • Tap for 2.4 mm Screws, Quick Coupling
  • Tap for 2.7 mm Screws, Quick Coupling
  • Graphics Aluminum Box with Silicone Fittings

Do you have any questions for 2.4 mm Locking L Plate Angle?
Please fill this form we will get in touch with you shortly.

Name(Required)
This field is for validation purposes and should be left unchanged.

Advantages of using locking plate for 2.4 mm Locking L Plate Angle

  • 2.4 mm Locking L Plate Angle is a locking plate so it does not have to precisely contact the underlying bone in all areas. When screws are tightened, they “lock” to the threaded screw holes of the plate, stabilizing the segments without pulling the bone to the plate. Locking screws make it impossible for screw insertion to alter the reduction. Nonlocking plate/screw systems require a precise adaptation of the plate to the underlying bone. Without this close contact, tightening of the screws will pull the bone segments toward the plate, resulting in loss of reduction and possibly the occlusal relationship
  • Locking plate/screw systems do not disrupt the underlying cortical bone perfusion as much as conventional plates, which compress the plate to the cortical bone.
  • Screws are unlikely to loosen from the plate. Similarly, if a bone graft is screwed to the plate, a locking head screw will not loosen during the phase of graft incorporation and healing. The possible advantage to this property of a locking plate/screw system is decreased risk of inflammatory complications due to hardware loosening.
  • Locking plate/screw systems have been shown to provide more stable fixation than conventional nonlocking plate/screw systems.

Locking Screw Technology

The heads of the locking screws contain male threads while the holes in the plates contain female threads. This allows the screw head to be threaded into the 2.4 mm Locking L Plate Angle hole, locking the screw into the plate. This technical innovation provides the ability to create a fixedangle construct while using familiar plating techniques.

Locking Plate Technology

By using locking screws in a bone plate, a fixed-angle construct is created. In osteopenic bone or fractures with multiple fragments, secure bone purchase with conventional screws may be compromised. Locking screws do not rely on bone/plate compression to resist patient load, but function similarly to multiple small angled blade plates. In osteopenic bone or comminuted fractures, the ability to lock screws into a fixed-angle construct is imperative.

By combining locking screw holes with compression screw slots in the shaft, the plate can be used as both a locking device and a fracture compression device. If compression is desired, it must be achieved first by inserting the standard screws in the compression screw slots before inserting any locking screws.

More Products from Hand and Foot Plates

Screws available for 2.4 mm Locking L Plate Angle

2 mm Locking Cortical Screws

2 mm Locking Cortical Screws available lengths are 8mm, 10mm, 12mm, 14mm, 16mm, 18mm, 20mm, 22mm, 24mm, 26mm, 28mm and 30mm.

2 mm Cortical Screws

2 mm Cortical Screws available lengths are 8mm, 10mm, 12mm, 14mm, 16mm, 18mm, 20mm, 22mm, 24mm, 26mm, 28mm and 30mm.

2.4 mm Locking Cortical Screws

2.4 mm Locking Cortical Screws available lengths are 8mm, 10mm, 12mm, 14mm, 16mm, 18mm, 20mm, 22mm, 24mm, 26mm, 28mm and 30mm.

2.4 mm Cortical Screws

2.4 mm Cortical Screws available lengths are 8mm, 10mm, 12mm, 14mm, 16mm, 18mm, 20mm, 22mm, 24mm, 26mm, 28mm and 30mm.

2.7 mm Locking Cortical Screws

2.7 mm Locking Cortical Screws available lengths are 8mm, 10mm, 12mm, 14mm, 16mm, 18mm, 20mm, 22mm, 24mm, 26mm, 28mm, 30mm, 32mm, 34mm, 36mm, 38mm and 40mm.

2.7 mm Cortical Screws

2.7 mm Cortical Screws available lengths are 8mm, 10mm, 12mm, 14mm, 16mm, 18mm, 20mm, 22mm, 24mm, 26mm, 28mm, 30mm, 32mm, 34mm, 36mm, 38mm and 40mm.

These Screws are made from pure Titanium and SS 316L. Any additional length sizes of these screws will be made on demand.

2.4 mm Locking L Plate Angle Surgical Technique

Select implant

Select the plates according to the fracture pattern and anatomy of the bone.

Reduce fracture

Reduce the fracture under image intensification and, if necessary, fix with Kirschner wires or reduction forceps. The reduction method is fracture-specific.

Trim 2.4 mm Locking L Plate Angle

Trim the plate to the desired length using bending/cutting pliers. Remove the burrs.

Contour plate

If necessary, contour the plate to suit the anatomical conditions. Use bending pins for LCP Plates and thread the pins into the screw holes to contour the plate. The plate can also be bent using two flat nosed pliers.

If possible, bend the plate between the locking/LCP holes. Do not deform the threaded part of the holes or overbend the plate as this may adversely affect insertion of locking screws. Avoid repetitive bending of the plate. Reverse bending or use of the incorrect instrumentation for bending may weaken the plate and lead to premature plate failure (e.g. breakage). Do not bend the plate beyond what is required to match the anatomy.

Position plate

Position the plate over the reduced fracture and, if necessary fix provisionally with Kirschner wires or reduction forceps. When using the holding forceps with swivel foot to hold the plate, remove the drill sleeves from the surrounding holes so that the holding forceps sit flush on the plate.

Determine screw type

Depending on the individual case, cortex and/or locking screws may be inserted. Determine where locking screws and where cortex screws will be used.
The final screw placement and the use of locking and cortex screws are determined by the fracture pattern. If a locking screw is inserted first, ensure that the 2.4 mm Locking L Plate Angle is held securely to the bone to prevent the plate from spinning as the screw locks into the plate.

Screw Insertion Cortical Screws

Drill the screw hole using the drill bit and the corresponding drill guide / drill sleeve. Determine the screw length with the corresponding depth
gauge. Insert the self-tapping cortex screw using the corresponding screwdriver.

Screw Insertion Locking Screws

Drill screw hole through the LCP drill sleeve using the appropriately sized drill bit. Determine the screw length either with the corresponding depth gauge or with the LCP drill sleeve. When using the LCP drill sleeve check the length directly on the scale of the drill sleeve. Then remove the drill sleeve. Insert the self-tapping locking screw using the corresponding screwdriver shaft and the appropriate handle.

Implant Removal

To remove the plate, first unlock all screws then definitely remove them in a second step. If the screws are not unlocked before removal the plate may rotate while the last screw is being removed and cause soft tissue damage.

2.4 mm Locking L Plate Angle Contraindications

Contraindications may be qualified or total, and need to be taken into consideration when evaluating the prognosis in each case. The physician’s education, training and professional judgement must be relied upon to choose the most appropriate device and treatment. Conditions presenting an increased risk of failure include:

  • Any active or suspected latent infection or marked local inflammation in or about the affected area.
  • Compromised vascularity that would inhibit adequate blood supply to the fracture or the operative site.
  • Bone stock compromised by disease, infection or prior implantation that can not provide adequate support and/or fixation of the devices.
  • Material sensitivity, documented or suspected.
  • Obesity. An overweight or obese patient can produce loads on the implant that can lead to failure of the fixation of the device or to
    failure of the device itself.
  • Patients having inadequate tissue coverage over the operative site.
  • Implant utilization that would interfere with anatomical structures or physiological performance.
  • Any mental or neuromuscular disorder which would create an unacceptable risk of fixation failure or complications in postoperative care.
  • Other medical or surgical conditions which would preclude the potential benefit of surgery.
  • All associated diseases which could endanger the function and success of the 2.4 mm Locking L Plate Angle.

Warnings and Precautionary for 2.4 mm Locking L Plate Angle

Before using 2.4 mm Locking L Plate Angle, the surgeon and ancillary staff should study the safety information in these instructions, as well as any product-specific information in the product description, surgical procedures and/or brochures.

Plates are made from medical grade materials and are designed, constructed and produced with utmost care. These quality assure best working results provided they are used in the proper manner. Therefore, the following instructions for use and safety recommendations must be observed.

Improper use of Plates can lead to damage to the tissue, premature wear, destruction of the instruments and injury to the operator, patients or other persons.

It is vital for the operating surgeon to take an active role in the medical management of their patients. The surgeon should thoroughly understand all aspects of the surgical procedure and instruments including their limitations. Care in appropriate selection and proper use of surgical instruments is the responsibility of the surgeon and the surgical team. Adequate surgical training should be completed before use of this plate.

Factors which could impair the success of the operation:

  • Allergies to implanted materials.
  • Localized bone tumours.
  • Osteoporosis or osteomalacia.
  • System disease and metabolic disturbances.
  • Alcohol and drug abuse.
  • Physical activities involving excessive shocks, whereby the implant is exposed to blows and/or excessive loading.
  • Patients who are mentally unable to understand and comply with the doctor’s instructions.
  • Poor general health.

Possible Adverse Effects

The following adverse effects are the most common resulting from implantation:

  • Loosening of the 2.4 mm Locking L Plate Angle, which may result from cyclic loading of the fixation site and/or tissue reaction of the implant.
  • Early and late infection.
  • Further bone fracture resulting from unusual stress or weakened bone substance.
  • Temporary or chronic neural damage resulting from pressure or hematomata.
  • Wound hematomas and delayed wound healing.
  • Vascular disease including venal thrombosis, pulmonary embolism and cardiac arrest.
  • Heterotopic ossification.
  • Pain and discomfort due to presence of the Implants.
  • Mechanical failure of the implant, including bending, loosening or breakage.
  • Migration of implant resulting in injury.

Preoperative Planning for 2.4 mm Locking L Plate Angle

The operating planning is carried out following a thorough clinical evaluation of the patient, Also, x-rays must be taken to allow a clear indication of the bony anatomy and associated deformities. At the time of the operation, the corresponding implantation instruments in addition to a complete set of 2.4 mm Locking L Plate Angle must be available.

The clinician should discuss with the patient the possible risks and complications associated with the use of Implants. It is important to determine pre-operatively whether the patient is allergic to any of the implant materials. Also, the patient needs to be informed that the performance of the device cannot be guaranteed as complications can affect the life expectancy of the device.

2.4 mm Locking L Plate Angle Precautions

  • Confirm functionality of instruments and check for wear during reprocessing. Replace worn or damaged instruments prior to use.
  • It is recommended to use the instruments identified for this screw.
  • Handle devices with care and dispose worn bone cutting instruments in a sharps container.
  • Always irrigate and apply suction for removal of debris potentially generated during implantation or removal.

2.4 mm Locking L Plate Angle Warnings

  • 2.4 mm Locking L Plate Angle can break during use (when subjected to excessive forces). While the surgeon must make the final decision on removal of the broken part based on associated risk in doing so, we recommend that whenever possible and practical for the individual patient, the broken part should be removed. Be aware that implants are not as strong as native bone. Implants subjected to substantial loads may fail.
  • Instruments, screws and cut plates may have sharp edges or moving joints that may pinch or tear user’s glove or skin.
  • Take care to remove all fragments that are not fixated during the surgery.
  • While the surgeon must make the final decision on implant removal, we recommend that whenever possible and practical for the individual patient, fixation devices should be removed once their service as an aid to healing is accomplished. Implant removal should be followed by adequate post-operative management to avoid refracture.

2.4 mm Locking L Plate Angle General Adverse Events

As with all major surgical procedures, risks, side effects and adverse events can occur. While many possible reactions may occur, some of the most common include: Problems resulting from anesthesia and patient positioning (e.g. nausea, vomiting, dental injuries, neurological impairments, etc.), thrombosis, embolism, infection, nerve and/or tooth root damage or injury of other critical structures including blood vessels, excessive bleeding, damage to soft tissues incl. swelling, abnormal scar formation, functional impairment of the musculoskeletal system, pain, discomfort or abnormal sensation due to the presence of the device, allergy or hypersensitivity reactions, side effects associated with hardware prominence, loosening, bending, or breakage of the device, mal-union, non-union or delayed union which may lead to breakage of the implant, reoperation.